UP-8000 Multi-Parameter Patient Monitor






Medical Equipment » Patient Monitor » UP-8000 Multi-Parameter Monitor

Applications
 
Our UP-8000 multi-parameter monitor is used for monitoring standard parameters of ECG, SpO2, RESP, NIBP, TEMP and PR. With a CSM module, it can also monitor parameters of CSI, EEG, EMG and BS% with alarms. The patient monitor is especially designed for anaesthesia use.
Features
 
1. The UP-8000 multi-parameter monitor uses a 12.1" high resolution color LCD screen.
2. It has a portable handset with wireless connection to main unit for data processing.
3. The multi-parameter monitor offers optimized NIBP measurement modes for adult, neonatal and pediatric patients.
4. It achieves automatic analysis of 20 arrhythmia waveforms.
5. Both automatic and manual analysis is available for S-T segment.
6. The patient monitor adopts digital SpO2 technology that helps ensure anti-motion and anti-low perfusion.
7. It can store up to 400 groups of data as well as 24 hour ECG waveforms.
8. The multi-parameter monitor offers protection against interference from defibrillator and electrosurgical generator.
9. It has a built-in thermal printer.

Specifications of UP-8000 Multi-Parameter Monitor
CSM


EEG sensitivity


±400μV
Noise level
<2μVp-p, <0.4μV RMS, 1~250Hz
CMR
>140dB
Input impedance
>50Mohm
Sample rate
2000 samples/sec. (14 bits equivalent)
CSI and update
0-100. Filter 6-42Hz, 1sec. update
EMG
0-100 logarithmic. Filter 75-85Hz, 1 sec. update
BS%
0-100%. Filter 2-42Hz, 1 sec. update


Wireless range


Up to 10 meters
ECG
Heart rate display range
15bpm ~ 350bpm
Heart rate display accuracy
± 1% or ± 2bpm, whichever is greater
Sweeping speed
12.5mm/s, 25mm/s, 50mm/s
Sensitivity selection
×1/2, 5mm/mV ×1, 10mm/mV ×2, 20mm/mV
CMRR
89dB
RESP


Measuring range


0rpm~120rpm
Accuracy
±5% or ±2 rpm
Alarm limits setting range
0rpm~120rpm
TEMP


Measuring range


25.0 ~45.0
Measuring accuracy
±0.2
Responding time
150s
NIBP
Measuring method
Oscillometric Technique
Measuring range
0mmHg~300mmHg
Accuracy
Maximal mean difference
±5 mmHg
Maximal standard deviation
8 mmHg
Measurement range
Adult


SYS 40 ~ 255 mmHg
MAP 20 ~ 215 mmHg
DIA 10 ~ 195 mmHg
Pediatric patient


SYS 40 ~ 200 mmHg
MAP 20 ~ 165 mmHg
DIA 10 ~ 150 mmHg
Neonate

SYS 40 ~ 135 mmHg
MAP 20 ~ 110 mmHg
DIA 10 ~ 95 mmHg
SpO2


Measuring range


35%~100%
Low perfusion capability
0.4%~5%
Accuracy
Not greater than 3% for SpO2range from 70% to 100%
PR


Measuring range


30bpm~240bpm
Measuring accuracy
±2bpm or ±2%, whichever is greater
General


Power supply


100~240VAC, 50/60Hz
Rechargeable Li-ion battery
12V 2.3Ah
Display mode
12 inches TFT color LCD
Alarming mode
Audible and visible alarm

Heal Force is an experienced multi-parameter monitor manufacturer and supplier in China. We provide various types of products such as electrosurgical generator, OR ceiling supply unit, CO2 incubator, and fetal Doppler monitor.